5 • Operation
Doc. 065031-04 1/08 147
5.5.1 Saving Manual Data
NOTE Chromeleon Xpress does not allow data to be saved.
If you are using Chromeleon, data from manual processing is saved in the
manual sequence folder in the timebase folder of the local datasource.
This data is overwritten each time a new manual sample is processed.
To save the data from a manual run:
1. Select the manual folder and select Save As from the File menu.
2. Enter a new name for the sequence and select the Save raw data
check box.
3. Click Save.
5.6 Automatic (Batch) Sample Processing
You can use Chromeleon or Chromeleon Xpress to create a list of samples (a
sequence) to be processed automatically.
• For each sample, the sequence includes a program with commands and
parameters for controlling ICS-3000 modules and acquiring sample data. To
create a program, click Create Program on the Sequence Control panel.
• If you are using Chromeleon, the sequence also includes a quantification
method for peak identification and area determination. To create a
quantification method, select New on the File menu and select Method File
from the list.
• You can also enter additional sample processing parameters in the sequence
(for example, the sample name, sample type, injection volume, etc.).
After creating the sequence, you can start batch processing.