EasyManua.ls Logo

Dionex ICS-3000 - DC Troubleshooting; Troubleshooting DC Error Messages

Dionex ICS-3000
392 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Doc. 065031-04 1/08 213
DC Troubleshooting
8.25 Troubleshooting DC Error Messages
If any of the following alarm conditions occurs, a message is displayed in the
Chromeleon or Chromeleon Xpress Audit Trail.
This error occurs when the conductivity cell is disconnected.
To troubleshoot:
1. Check the detector connection: An electronics connector on the back of
the detector plugs into a receptacle on the detector compartment. Push on
the upper part of the detector (see Figure 8-3
) to make certain the
connector is securely connected.
2. If the error persists, there may be a problem in the detector electronics.
Contact Dionex for assistance.
NOTE The ICS-3000 Ion Chromatography System
electronics components cannot be serviced by the
user.
CD cell 1 option disconnected.
-or-
CD cell 1 option disconnected.
Figure 8-3. Conductivity Detector Connection
Push here to ensure the detectors
electrical connection is secure.

Table of Contents

Related product manuals