ICS-3000 Ion Chromatography System
214 Doc. 065031-04 1/08
This error occurs when the temperature of the specified conductivity detector
is higher than the maximum allowed. This error may occur if the ICS-3000
Ion Chromatography System is operating in an environment in which the
temperature is greater than 40 °C (104 °F).
To troubleshoot:
1. Refer to Section A.12
for environmental specifications.
2. Verify that the compartment setpoint is at least 5 °C less than the CD cell
setpoint.
3. Check the suppressor current setting. Running the suppressor at a higher
current than is recommended for the application can cause heat up of the
CD cell.
This error may indicate a problem in the detector electronics. Contact Dionex
for assistance.
This error occurs when the temperature of the column compartment exceeds
the maximum allowed. This error may occur if the ICS-3000 Ion
Chromatography System is operating in an environment in which the
temperature is greater than 40 °C (104 °F).
To troubleshoot:
Refer to Section A.12
for environmental specifications.
CD cell 1 over safe temperature.
-or-
CD cell 2 over safe temperature.
Column open circuit.
Column over safe temperature.