8 • DC Troubleshooting
Doc. 065031-04 1/08 221
To troubleshoot:
1. Verify that the suppressor cable is connected correctly.
This message occurs when the door to the upper compartment is opened
during a run.
To troubleshoot:
1. Verify that the door is fully closed.
2. Check for—and remove—any obstruction.
3. If the door is fully closed and the error persists, contact Dionex for
assistance.
8.26 Liquid Leaks from DC Components
• Leaking fitting
Locate the source of the leak. Tighten or, if necessary, replace the liquid line
connection (see Section 9.17
).
• Broken liquid line
Replace the line and fittings with the same length and internal diameter tubing
(see Section 9.17
).
• Blocked or improperly installed line
Make sure the lines are not crimped or otherwise blocked. Also, if the blocked
line is a waste line, make sure it is not elevated at any point after it exits the
DC. If a line is blocked, replace it (see Section 9.17
).
Suppressor 1 open circuit.
-or-
Suppressor 2 open circuit.
Upper door opened.