TC Troubleshooting
Doc. 065031-04 1/08 255
8.33 TC Temperature Stabilizer Not Operating Properly
Temperature stabilizer installed incorrectly
Remove the temperature stabilizer and reinstall it correctly. Make sure there is
good thermal contact between the temperature stabilizer and the thermal
element. For detailed instructions, refer to the ICS-3000 Ion Chromatography
System Installation Instructions (Document No. 065032).
TC calibration is incorrect
The TC cannot be calibrated by the user. Contact Dionex for assistance.
Defective temperature stabilizer
Replace the temperature stabilizer (standard bore, P/N 064548; microbore,
P/N 064650).