Doc. 065031-04 1/08 151
6 Shutdown
DP/SP Shutdown
If the ICS-3000 Dual Pump (DP) or ICS-3000 Single Pump (SP) will not be
operated for a period of one week or more, follow the instructions below:
Fill the pump with methanol (or a similar alcohol, such as 2-propanol or
ethanol). If the eluents in the pump are not miscible with water, replace the
eluents step-by-step.
Fill the seal wash reservoir with wash solution (see Section 2.2.5).
Rinse out buffers. This will reduce the time required for column equilibration
when you resume operation.
Disconnect the tubing from the peristaltic pump on the component panel: lift
the lever up and to the right, remove the tubing, and release the lever (see
Figure 6-1
).
Figure 6-1. Peristaltic Pump for Seal Wash System
Rotor
Lever
Tubing