EasyManua.ls Logo

Dionex ICS-3000 - DC Service; Replacing Tubing and Fittings

Dionex ICS-3000
392 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Doc. 065031-04 1/08 297
DC Service
9.17 Replacing Tubing and Fittings
The DC is plumbed with the tubing and fittings listed below.
Use 1/8-in fittings (P/N 052267) and ferrules (P/N 048949) for connections to
the suppressor
REGEN OUT port and the eluent reservoir.
Use double-cone 10-32 ferrule fittings (P/N 043276) and 10-32 bolts
(P/N 043275) for all other tubing connections. For tightening requirements,
refer to Installation of Dionex Liquid Line Fittings (Document No. 031432).
The manual is on the Dionex Reference Library CD-ROM (P/N 053891).
Tubing Size and Type Color P/N Used For
0.75-mm (0.030-in) ID
PEEK
Green 044777 Connection from injection
valve to waste
AS needle seal assembly,
46 cm (18 in), with 10-32
ferrule fitting
Pink 057301 AS normal and concentrate
modes only: Connection from
the injection valve to the AS
inject port
0.33-mm (0.013-in) ID,
PEEK
Light
Blue
049714 AS sequential modes only:
Connection from injection
valves to AS diverter valve
0.51-mm (0.020-in) ID,
PEEK
Orange 042855 AS simultaneous mode only:
Connection from each injection
valve to the AS simultaneous
injection tee
1.58-mm (0.062-in) ID
Teflon
Colorless 014157
Connection from
REGEN OUT
port on the suppressor
0.125-mm (0.005-in) ID
PEEK
Red 044221 Microbore systems:
Connections between other
system components
0.25-mm (0.010-in) ID
PEEK
Black 042690 Standard bore systems:
Connections between other
system components

Table of Contents