ICS-3000 Ion Chromatography System
294 Doc. 065031-04 1/08
9.14 Replacing the RFIC Eluent Degasser
The RFIC Eluent Degasser contains a tubing assembly that may rupture if
subjected to excessive pressure.
1. Press the
POWER button on the front of the EG to turn off the power.
2. Open the front door of the EG.
3. Push down on the slide release latch and pull the tray forward until it reaches
the stop.
4. Grasp the degasser by the sides and pull it straight off the ball studs that
secure it to the component mounting panel. Dispose of the degasser.
5. Disconnect the four lines connected to the RFIC Eluent Degasser fittings
(
ELUENT OUT, REGEN OUT, REGEN IN, and ELUENT IN).
6. Connect the four lines that were disconnected in Step 5
to the new RFIC
Eluent Degasser (P/N 062137).
7. Align the two mounting holes on the back plate of the new RFIC Eluent
Degasser with the ball studs on the component mounting panel. Push the
degasser firmly onto the mounting ball studs. The degasser clicks into place
when properly installed.
8. Push the tray into the module until it locks into place. Close the EG door.
9. Turn on the power and resume operation.