4• DP/SP Startup
Doc. 065031-04 1/08 127
b. In the Commands dialog box, select the pump name.
c. Select the RearSealWashSystem command and select an option:
• Select Internal if the standard seal wash reservoir is installed.
• Select Automatic if the drop-sensor cap reservoir is installed.
d. Click Execute.
4.6 Start the Pump
1. Press the POWER button on the front of the DP/SP.
2. If any of the following conditions applies, prime the pump (see Section 9.3
)
before proceeding:
• The eluent has been changed.
• The eluent line is new (empty).
• The eluent line contains air.
3. To display the ICS-3000 panel tabset, follow one of the steps below:
• If Chromeleon is installed, launch the application and click the Default
Panel Tabset toolbar button.
• If Chromeleon Xpress is installed, launch the application; this
automatically displays the ICS-3000 panel tabset.
4. The panel tabset opens to the Home panel by default. This panel displays
basic status information for each instrument in the system. In addition, a