7 Maintenance
This section describes routine maintenance procedures for the ICS-3000 system
that users may perform. All other maintenance procedures must be performed by
Dionex personnel.
For information about maintenance procedures for the EluGen cartridge, CR-TC
trap column, or suppressor, refer to the appropriate product manual. These
manuals are included on the Dionex Reference Library CD-ROM (P/N 053891).
7.1 System Maintenance Checklists
7.1.1 Daily Maintenance
Check for leaks.
Wipe up liquid spills.
Check eluent reservoir levels and refill as needed.
Check eluent reservoirs for microbial growth; clean reservoirs and
replace eluent as needed.
Check waste containers and empty as needed.
For detailed daily maintenance instructions for each module, see:
7.1.2 Weekly Maintenance
Check end-line filters on eluent lines. When end-line filters are new,
they are pure white. Replace filters if they become discolored, if
bacterial buildup is evident, or if eluent flow is blocked. See
Section 4.4 for instructions.
Check for plugged or blocked tubing connections.
DP/SP page 161
EG page 163
DC page 165
TC page 167