End-line filters are provided in the ship kit (DP Ship Kit, P/N 062463; SP
Ship Kit, P/N 063342). Before installation, thoroughly rinse the end-line filter
with ASTM Type I (18 megohm-cm) filtered and deionized water to remove
any loose particles.
• Replace the seal wash solution. ASTM Type I (18 megohm-cm) filtered and
deionized water is appropriate for most applications.
• Check the tubing connected to the peristaltic pump. If the tubing is blocked or
crimped, replace it (see Section 9.7
).
7.4 DP/SP Periodic Maintenance
• Make fresh eluent as needed.
• Replace piston seals every 6 to 12 months (see Section 9.5).
7.5 DP/SP Annual Maintenance
• Dionex recommends performing preventive maintenance annually. The
following kits are available:
DP Preventive Maintenance Kit (P/N 061794)
SP Preventive Maintenance Kit (P/N 061795)