ICS-3000 Ion Chromatography System
162 Doc. 065031-04 1/08
End-line filters are provided in the ship kit (DP Ship Kit, P/N 062463; SP
Ship Kit, P/N 063342). Before installation, thoroughly rinse the end-line filter
with ASTM Type I (18 megohm-cm) filtered and deionized water to remove
any loose particles.
Replace the seal wash solution. ASTM Type I (18 megohm-cm) filtered and
deionized water is appropriate for most applications.
Check the tubing connected to the peristaltic pump. If the tubing is blocked or
crimped, replace it (see Section 9.7
).
7.4 DP/SP Periodic Maintenance
Make fresh eluent as needed.
Replace piston seals every 6 to 12 months (see Section 9.5).
7.5 DP/SP Annual Maintenance
Dionex recommends performing preventive maintenance annually. The
following kits are available:
DP Preventive Maintenance Kit (P/N 061794)
SP Preventive Maintenance Kit (P/N 061795)