ICS-3000 Ion Chromatography System
220 Doc. 065031-04 1/08
3. If you suspect a malfunction in the suppressor controller, contact Dionex
for assistance.
NOTE The ICS-3000 Ion Chromatography System electronics
components cannot be serviced by the user.
This error appears when, in order to maintain the selected current, the ICS-
3000 Ion Chromatography System is required to apply a higher voltage than
the suppressor can support.
To troubleshoot:
1. Reduce the flow rate.
2. Rehydrate the suppressor. Refer to the suppressor manual for instructions.
Suppressor manuals are included on the Dionex Reference Library CD-
ROM (P/N 053891).
3. If the error persists, replace the suppressor (see Section 9.23.2
).
This error appears if you turn on the suppressor and the system cannot
establish a connection with the suppressor.
To troubleshoot:
1. Check the suppressor cable connection.
2. If the error persists, replace the suppressor (see Section 9.23.2
).
Suppressor 1 over-power.
-or-
Suppressor 2 over-power.
Suppressor 1 over-voltage.
-or-
Suppressor 2 over-voltage.