ICS-3000 Ion Chromatography System
132 Doc. 065031-04 1/08
Single-Cartridge or Independent Dual-Cartridge Configuration
In the single-cartridge configuration, the EG contains one EluGen cartridge. In the
independent dual-cartridge configuration, the EG contains two EluGen cartridges
operating independently on separate systems (each cartridge is linked to a
different DP/SP). For details, refer to the EluGen cartridge manual. Cartridge
manuals are included on the Dionex Reference Library CD-ROM (P/N 053891).
EluGen Cartridge Eluent Concentration Range
K
2
CO
3
0.1 to 15 mM at 0.1to 1.0 mL/min flow
0.1 to X mM at 1.0 to 2.0 mL/min flow
where X = 15/flow
KOH 0.1 to 100 mM at 0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 100/flow
LiOH 0.1 to 80 mM at 0.1 to ≤ 1.0 mL/min flow
0.1 to X mM at 1.0 to ≤ 3.0 mL/min flow
where X = 80/flow
MSA 0.1 to 100 mM at 0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 100/flow
NaOH 0.1 to 100 mM at 0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 100/flow
Table 4-1. Eluent Concentration Ranges for Single-Cartridge and
Independent Dual-Cartridge Configurations