ICS-3000 Ion Chromatography System
36 Doc. 065031-04 1/08
Electrical Connections
Electrical connectors provide a connection from the components installed in the
EG to the EG power supplies. There are two sets of electrical connectors (one per
channel). Connections for the following components are provided:
Two EluGen cartridges or one cartridge and one EPM Electrolytic pH
Modifier
and
Two Continuously Regenerated Trap Columns (CR-TC)
When operating with an EGC II K
2
CO
3
EluGen Cartridge and the EPM, the
cartridge is connected to the EGC power supply for one channel and the EPM is
connected to the EGC power supply for the second channel. Thus, the EG can
accommodate only one such configuration.
Waste Line
The module waste line (white tubing) discharges liquid and gas waste from the
RFIC Eluent Degasser. This line is marked
WASTE-GAS SEPARATOR inside the
EG.
NOTE A gas separator waste tube (P/N 045460) should be
connected to the EG waste line during installation. For
details, refer to the ICS-3000 Ion Chromatography
System Installation Instructions (Document No. 065032).
Place the free end of the waste line into a waste container. To maintain an active
drain, position the waste container below the level of the EG.
NOTE To ensure proper drainage, each waste line must remain
above the liquid level in the waste container.