ICS-3000 Ion Chromatography System
48 Doc. 065031-04 1/08
Switched AC Sockets
The two AC receptacles can be used to control the power to external devices. Use
Chromeleon or Chromeleon Xpress to switch the power on and off. TTL inputs
can also be used to control the AC sockets (see Section 2.18.4
).
Fuse Holder, Main Power Switch, and Power Receptacle
The fuse holder contains two 10-amp slow-blow fuses (P/N 954772). For
instructions on how to change the fuses, see Section 9.19
.
The rear panel power switch is the main power switch for the DC. Turn on the
main power switch before initial operation and leave the switch on unless
instructed to turn it off (for example, before performing a service procedure).
NOTE For routine on/off control, use the POWER button on the
front of the DC (see Figure 2-12
). To turn off the DC,
press and hold the
POWER button for 2 seconds.
The power cord plugs into the IEC 320 three-prong receptacle.
USB Connections
The USB (Universal Serial Bus) receptacle (“B” type connector) allows
connection to the PC on which Chromeleon or Chromeleon Xpress software is
installed.
The two USB ports (“A” type connectors) allow connection to other USB devices
in the system (for example, the pump and eluent generator).
A 1.8-m (6-ft) USB cable (P/N 960777) is included in the DC Ship Kit
(P/N 062614, dual; P/N 063408, single).
The power supply cord is used as the main disconnect device. Make
sure the socket-outlet is located near the DC and is easily accessible.
Le cordon d'alimentation principal est utilisé comme dispositif
principal de débranchement. Veillez à ce que la prise de base soit
située/installée près du module et facilement accessible.
Das Netzkabel ist das wichtigste Mittel zur Stromunterbrechung.
Stellen Sie sicher, daß sich die Steckdose nahe am Gerät befindet und
leicht zugänglich ist.