An additional trigger will display on the ECG trace.
3. To delete a trigger point on the ECG trace and to normalize the heart cycle, place the cursor
on the trigger and right-click
Delete R-peak Trigger
.
4.
To move a trigger, place the cursor over the trigger and click and drag it to the new location.
6.8 Change heartbeat from Arrhythmia Retrigger to Reconstruct
1. Place the cursor in the ECG Editor, and right-click and select
Change Scan to Reconstruct
.
2.
The reconstruction window moves from one heartbeat to the other heartbeat.
6.9 Display secondary recon images
6.9.1 Cardiac image annotation
SSCIN (Secondary Cardiac Axial — single sector image)
6.9.2 Edited cardiac image annotation
Image where the ECG trace has been edited, are annotated with an “E” placed before the scan
type, e.g. E/SSCIN.
E/SSCIN (Secondary Cardiac Axial — single sector image)
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
366 6 Cardiac Recon