EasyManua.ls Logo

GE Revolution CT - Graphically Adjust Reconstruction Timing; Method 1; Method 1 and 2; Method 2

GE Revolution CT
702 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Number Description
2 Secondary reconstructions parameters are displayed: start phase, end phase, and interval.
3
The trigger point on the ECG Trace for each R-peak is displayed, as well as the X-ray on interval, and
the image reconstruction timing when images will be generated.
4 Click the [Save] icon to save your modified trigger locations with exam information.
5 Click the [Restore] icon to restore the original ECG Trace information for the acquisition.
6
Click the [Measure] icon to measure, in msec, the distance between two defined points on the wave‐
form. Place the cursor on the trace and click and drag right or left.
7 Click the [Magnify] icon or the [Minimize] icon to magnify or minimize the ECG waveform.
8 Click and drag the navigation bar to view the selected portion of the waveform.
6.6 Graphically adjust reconstruction timing
Use this procedure to move or reposition image reconstruction timing within a single heart cycle.
6.6.1 Method 1
Place the cursor over the recon location and click, drag, and drop to the new location.
The annotation in the reconstruction location indicates the % or ms where images will be
reconstructed.
6.6.2 Method 2
1. Place the cursor over the recon window you wish to move, and right-click
Adjust Recon
Timing
.
2.
Type and enter the location as + or - milliseconds.
The recon timing for selected heart cycle moves to the new location.
3. Click [OK].
6.6.3 Method 1 and 2
For all cardiac scan types, the recon window cannot be moved outside of its current X-ray on
interval.
6.7 Insert, delete, or move an R-peak trigger
Use this procedure if gating issues occur during the scan or if the system does not trigger
accurately at the R-peak of the ECG trace.
Insert, remove, or move a trigger to normalize a heart cycle when a trigger occurs at an
unwanted location. This can be due to abnormal ECG waveform patterns, external noise
interference, or a low amplitude trace.
Remove an image reconstruction location from one or more heart cycles to improve image
quality when an arrhythmia is present in the scan.
1. Open a prior ECG-gated scan series in the
Reconstruction and Image Processing
area on
the image monitor.
2. To insert a trigger, place the cursor anywhere on the ECG trace where you want to insert it
and right-click
Insert R-peak Trigger
.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Chapter 13 Cardiac 365

Table of Contents

Related product manuals