EasyManuals Logo

GE Revolution CT User Manual

GE Revolution CT
702 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #291 background imageLoading...
Page #291 background image
NOTE: Flip/Rotate options are only available after service checks that the viewing stations
will display the data correctly.
6.1.9 Copy Forward
Copy Forward Patient Orientation, Patient Position and Anatomical Reference is a feature that
adjusts these three bundled parameters in a scanning protocol based on the values in the
previous series. This feature is located in Protocol Management.
NOTE:
Copy Forward is set at the destination or recipient series.
For example, if Copy Forward was turned “On” in Series 2 of a protocol, and during
scan Patient Orientation, Patient Position and Anatomical Reference were changed
from Head First, Supine, SN to Feet First, Prone, OM for Series 1, these three
settings will be automatically copied for scanning Series 2.
Similar to Copy Patient Orientation / Position, Copy Forward works within the same
patient exam.
The Copy Forward feature can also be used to duplicate (reuse) the following parameters within
series from primary reconstruction to secondary reconstructions:
Start Location
End Location
Time Between Images (Cine scan type only)
Interval (Helical scan type only)
DFOV
R/L Center
A/P Center
To activate Copy Forward, type d or D (duplicate) in the
Parameter
text field in the receiving
reconstruction.
6.2 Primary Recon collection
From the
Scan Settings
window, click the
Primary Recon
collection.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Chapter 11 Scan 291

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Revolution CT and is the answer not in the manual?

GE Revolution CT Specifications

General IconGeneral
Detector CoverageUp to 160 mm
Gantry Tilt± 30 degrees
Spatial ResolutionUp to 0.23 mm
Scan RangeUp to 200 cm
X-ray Tube CurrentUp to 800 mA
TypeComputed Tomography (CT) Scanner
Rotation Speed0.28 seconds per rotation
Tube Voltage80 kV to 140 kV
Number of Detector Rows256
Power Requirements480 VAC, 3-phase
Detector Rows256
Coverage160mm

Related product manuals