3.3.8 Duplicate a series
Use this procedure to duplicate a series that has already been scanned.
On the
Scan Settings
window, right-click on the series to be duplicated in the task list and select
Duplicate Series
.
NOTE: The Prep Delay for group 1 is set to zero.
3.3.9 Stop a scan
3.3.9.1 Abort a scan in progress
Use this procedure to abort the X-ray and stop gantry and table movement.
1.
On the scan control interface, press <Stop Scan>.
2. Click <Resume> from the
Scan Progress
window to resume the scan.
3.3.9.2 Pause a scan in progress
Use this procedure to pause the current scan. This finishes the current scan, then pauses the
next scan.
1.
On the scan control interface, press <Pause Scan>.
2. Click [Resume] from the
Scan Progress
window to resume the scan.
3.3.9.3 Emergency stop scan, electronics, and x-radiation
Use this procedure in the event of a patient related emergency or if the cradle, table, or gantry
starts to move unexpectedly.
1.
On the scan control interface or gantry, press <Emergency Stop>.
All table and gantry motions and X-ray exposures are stopped.
2.
On the gantry, press <Reset> to clear the Emergency Stop.
NOTE: For helical scans, you need to evaluate each situation. <Pause Scan> will not stop a
helical scan in progress. The scan will complete and the system will stop and not scan
any additional groups. <Stop Scan> will stop a helical scan. [Resume] will resume a
helical scan from the location it was stopped.
3.3.10 Close the exam
When an exam is completed, click [Done Scanning], then [Close Exam].
•
[Done Scanning] transitions the exam from a scanning session (on the scan monitor) to
image processing done scanning session (on the image monitor).
•
[Close Exam] removes the scanning session completely from the user interface.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
264 3 Exam Workflow