Table 16: Coverage Speed — Cine
Element Description
Table Positions One
(Single)
Detector Coverage
•
40 mm
•
80 mm
•
120 mm
•
140 mm
•
160 mm
Also see the
Equipment
chapter;
Detector Configuration
section for more information.
Number of Passes
The number of rotations for the data acquisition.
Duration
1–60 s
Scan Interval
0
Rotation Time
0.28 s, 0.35 s, 0.5 s, 1.0 s
Total Exposure Time
Determined by the duration.
4.3 kV and mA Control
From the
Scan Settings
window, click
kV and mA Control
to open the
kV and mA Control
collection.
Settings include:
kV Mode
,
kV
,
mA Mode
and
Noise Index
.
Illustration 16: kV and mA Control collection
4.3.1 kV Mode
From the
kV and mA Control
collection, select
kV Mode
and/or
kV
.
4.3.1.1 Manual kV Mode
When
Manual
kV Mode is selected, you can enter a kVp value for each prescribed scan group.
When building protocols, make sure that a reasonable and appropriate kVp level is selected for
the Manual kV selection even if the protocol is designed to employ kV Assist. This is done to
ensure that a reasonable kVp setting is used for the diagnostic scan in the event that kV Assist
is disabled.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
270 4 Scan Parameters