8 Build Protocols with Notification Values by Factor
Use this procedure to set the Notification Values (NV) in user protocols for all scan groups by a
factor.
For example, if the CTDI
vol
in a scan group were 5 and a factor of 1.5 was entered, the new NV
value for the group would be 7.5.
•
Values can by entered in increments of 0.1.
•
Values must be > 0.
•
The maximum value is 10.
1.
From the image monitor, click
.
2. From the
Mode
menu, click
Protocol Management
, then
Protocol Management
.
3. Enter your Log On Name and Password and click [OK] to continue.
4.
From the
Protocol Selection
window, select
Adult
or
Pediatric
.
5. From the
Anatomical
area, select an anatomical region.
6. From the
Protocol Selection
window, select a protocol and click [Edit].
7. Click the series task to go to the first non-scout series in the protocol.
8. From the
Dose Display
area, click
NV CTDI
vol
or
DLP
.
9.
Select
Factor
from the drop-down menu and type the desired factor for
CTDI
vol
and/or
DLP
.
10. Click [Confirm Changes].
11. Click [Save] or [Save As] to return to the
Protocol Selection
window.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Chapter 7 Dose Check 149