HD Ultra
5. If Cardiac Hi Res mode is enabled, select a higher resolution cardiac algorithm that has the
same characteristics as the non-Hi Res cardiac algorithms.
HD Soft
HD Stnd
HD Stnd Plus
HD Detail
HD Detail Plus
HD Lung
HD Edge
6.3.4 Set Recon Options for Helical scan types
6.3.4.1 Recon Mode
To adjust the
Recon Mode
, select
Full
or
Plus
.
NOTE:
Plus Recon mode is not annotated on the Series Text Page.
Full mode is not allowed with 0.625 mm slice thickness at either 0.984:1 pitch or 1.375:1 pitch.
6.3.4.2 IQ Enhance
To adjust IQ Enhance, click
IQ Enhance
to select either
On
or
Off
.
IQ Enhance is used to minimize artifacts seen in helical thin slice images.
IQ Enhance is allowed with both 0.625 mm and 1.25 mm slice thickness, and the image interval
can either be 50% or 100% of the slice thickness.
IQ Enhance annotation is added to the left side of the image.
6.3.5 ASiR-V for all scan types
Follow these steps to enable ASiR-V and select an ASiR-V percentage level.
1.
From the
Primary Recon
collection, click
ASiR-V
.
2. Select the desired ASiR-V percentage level.
3. Click [Apply].
6.3.6 Adjust Window Width and Window Level for all scan types
Follow these steps to adjust Window Width and Window Level.
1.
Open the
Primary Recon
collection.
2. Click
Window Width
and type the desired value.
3. Click
Window Level
and type the desired value.