Image Optimization
5-46 Vivid S70 / S60 – User Manual
BC092760-1EN
01
Additional scanning features
LogiqView
LogiqView provides the ability to construct and view a static 2D
image which is wider than the field of view of a given transducer.
This feature allows viewing and measurements of anatomy that
is larger than what would fit in a single image.
LogiqView constructs the extended image from individual image
frames as the operator slides the transducer along the surface
of the skin in the direction of the scan plane.
LogiqView is available with all linear array probes.
Using LogiqView
1. Perform a detailed examination of the anatomy/pathology.
Optimize parameters for tissue texture and visible window
prior to activating LogiqView.
2. Press LogiqView.
3. To start acquiring the image, press 2D freeze key.
Scan slowly and in a uniform motion lengthwise.
• Continuous contact is required throughout the length of
the extended image.
• Always keep the transducer perpendicular to the skin
surface.
• Keep the motion within the same scan plane.
• Do not make abrupt changes in speed of motion.
4. If required, press 2D freeze again to restart LogicView.
5. To complete the scan, press Freeze.
6. Adjust LogiqView rotate to rotate the acquisition.
NOTE: The quality of the resulting image is somewhat user-dependent
and requires some additional skill and practice to develop
proper technique.