EasyManua.ls Logo

GE Vivid S60

GE Vivid S60
709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Prepare the System for Use
3-4 Vivid S70 / S60User Manual
BC092760-1EN
01
Connecting the system
A GE-qualified person should perform the initial system
installation.
Connecting the ultrasound system involves preliminary checks
of the power cord, voltage level, and compliance with electrical
safety requirements.
Use only power supply cords, cables, and plugs provided by or
designated by GE.
Ensure that the power cord and plug are intact and that the
power plug is the proper hospital-grade type (where required).
WARNING
The system should be connected to a fixed power socket which
has the protective grounding connector. Never use an
extension cord or adapter plug.
WARNING
Failure to provide an adequate earth circuit can cause electrical
shock, resulting in serious injury.
WARNING
Connection of additional protective earth conductors or
potential equalization conductors is not necessary in most
cases and is only recommended for situations involving
multiple equipment in a high-risk patient environment to
provide assurance that all equipment is at the same potential
and operates within acceptable leakage current limits. An
example of a high-risk patient would be a special procedure
during which the patient has an accessible conductive path to
the heart such as exposed cardiac pacing leads.

Table of Contents

Other manuals for GE Vivid S60

Related product manuals