Connectivity configuration
Vivid S70 / S60 – User Manual 12-43
BC092760-1EN 01
• The IP address for the system
• The subnet mask for the system
• The IP address for the Default Gateway
4. Press Save settings and reboot the system.
System in a network environment with a DICOM server
The system’s TCP/IP settings must be configured as described
in ‘System’s TCP/IP settings’ on page 12-42.
In addition, to work against the DICOM server the following
information has to be entered in the system:
• The DICOM server IP address
• The DICOM server port number
• The DICOM server AE title (the server’s name)
NOTE: Changing the DICOM server AE title requires a reboot of the
system to take effect.
Typically source for this information is the network administrator.
Create or modify a DICOM server IP address
Follow the steps below if the IP address settings for the DICOM
server need to be modified or created:
1. Press Utility/Config on the control panel and log on as
administrator.
2. Select the Connectivity category and TCP/IP subgroup.
The TCP/IP sheet is displayed.
Figure 12-27. TCP/IP sheet - Server config section
3. In the Server config section, select the DICOM server and
press Modify (or press Add if creating a new IP address).
The Server config window is displayed.

Table of Contents

Other manuals for GE Vivid S60

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60 and is the answer not in the manual?

GE Vivid S60 Specifications

General IconGeneral
BrandGE
ModelVivid S60
CategoryMedical Equipment
LanguageEnglish

Related product manuals