M-Mode
Vivid S70 / S60 – User Manual 5-11
BC092760-1EN 01
Optimizing M-Mode
The use of preset gives optimum performance with minimum
adjustment. If necessary, the following controls can be adjusted
to further optimize the M-Mode display:
NOTE: Refer to ‘Generating a new preset’ on page 12-90 about
creating presets.
• Adjust Horizontal sweep to optimize the display resolution.
• Adjust Gain and TGC controls to adjust the range to be
imaged.
• Use the Frequency (move to higher frequencies) or the
Frame rate control (move to lower frame rate) to increase
resolution in image.
• Adjust Dynamic range to optimize the useful range of
incoming echoes to the available grayscale.
• Adjust Compress and Edge Enhance to further optimize
the display.
• Adjust Reject to reduce noise while taking care not to
eliminate significant low-level diagnostic information.
• Press Octave to toggle between fundamental and Harmonic
mode.
• Use the Focus Pos control to center the focal point around
the region of interest.
• Adjust Power to obtain an acceptable image using the
lowest setting possible.
NOTE: The Power setting affects all other operating modes.