01
Anatomical M-Mode
M-Mode applied to TVI, Tissue Tracking, Strain rate, Strain or
intensity data (Contrast) calculates and color/codes data
accordingly along a path drawn by the operator.
Using Anatomical M-Mode
1. Press Curved AMM.
2. In one of the Cineloop windows, place the first point of the
path.
3. Move the cursor to the location for the next anchoring point
of the path and press Select.
A path with two anchor points will give a straight anatomical
M-Mode profile. By creating more than two anchor points,
the user can bend the path and obtain a curved anatomical
M-Mode profile.
NOTE: To edit a path under construction, move the cursor
backward and retrace the path.
4. To end the trace press Select twice (double clicking).
The color-coded display of the corresponding data
calculated along the path is shown in the Analysis window.
NOTE: Adjust Horiz. Sweep and scroll through the cineloop to
optimize the display to the portion of interest.
Optimizing Anatomical M-Mode
Edition of the curve
The drawn Anatomical M-Mode path can be edited by moving
the anchor points.
To move an anchor point
1. Select the anchor point to move.
2. Move the anchor point to a new position.
3. Press Select to anchor the point to its new location.