Probes
13-8 Vivid S70 / S60 – User Manual
BC092760-1EN
01
Maximum probe temperature
NOTE: Lens temperature measured under following conditions per
IEC 60601-2-37 Ed2:
1. Thermocouple is placed at the maximum temperature spot
of the lens. Ambient temperature at 23 °C ±3 °C.
Probe
Max Temp (°C)
(Simulated use)
Max Temp (°C)
(Still air)
M5Sc-D (Vivid S70 only) 42.0 42.2
3Sc-RS 41.1 37.1
6S-D 41.3 39.9
12S-D 41.5 33.1
9L-D 41.6 37.6
11L-D 41.8 38.0
L8-18i-D 41.9 36.5
C1-5-D 41.1 40.5
C1-6-D 41.4 39.5
C2-9-D 38.6 37.3
C3-10-D 41.1 36.8
IC5-9-D 41.9 35.4
2D (P2D) 35.7 35.5
6D (P6D) 34.8 28.1
6VT-D 40.8 36.0
6Tc-RS 41.7 37.9
9T-RS 41.6 33.6
10T-D (Vivid S70 only) 41.9 42.5
AcuNav
TM
10F
37.0 34.9
AcuNav
TM
8F
37.8 33.6
SoundStar
TM
3D 10F
37.0 34.9
SoundStar
TM
eco 10F
37.0 34.9
SoundStar
TM
eco 8F
37.8 33.6