Maintenance
15-8 Vivid S70 / S60 – User Manual
BC092760-1EN
01
System self-test
System malfunction
The ultrasound system is designed for reliable operation and
consistent, high-quality performance. Automatic self-testing
facilities are provided to monitor system operation and to detect
malfunction as soon as possible, thereby eliminating
unnecessary downtime. The detection of any serious
malfunction may result in immediate interruption of the system
operation.
In the event of error or system malfunction the user may save
locally or export a log file to a removable media as described
below and contact authorized service personnel.
In addition, system malfunctions can be bookmarked, enabling
creation of a log file specific to that event.
Bookmarking a system malfunction
1. If a system malfunction is observed, press Alt - B.
A bookmark will be created when creating a log file.
Generating a log file
1. Press Alt - D on the alphanumeric keyboard.
The Problem description dialogue window is displayed (see
Figure 15-1).
2. Type in a description of the problem. Notes should be made
regarding the selected probe, the imaging mode and the
application that was being used at the time of malfunction. If
applicable, try to describe the button or key pushing
sequence that immediately preceded the problem.
Check the mention System lockup if applicable.
3. Select the destination where to save or export the log file.
If Store locally is selected, the log file is saved to the local
hard disk.
If a removable media is selected, the current and previously
saved log files are exported to the selected media.