12-42 Vivid S70 / S60User Manual
BC092760-1EN
01
Connectivity configuration
System in a network environment
To be able to use the network functions when connected to a
hospital network, the system must have a proper network
address. Typically source for this information is the network
administrator.
System’s TCP/IP settings
1. Press Utility/Config on the control panel and log on as
administrator.
2. Select the Connectivity category and TCP/IP subgroup.
The TCPIP subgroup is displayed.
Figure 12-26. TCP/IP subgroup
3. Select Network settings to configure:
1. Section My Computer
Computer name: not editable
IP address: press Network Settings to edit the IP address in Windows.
2. Section Server Config
Add, modify or remove server connections (EchoPAC Software Only, EchoServer, ImageVault), see
‘Create or modify a DICOM server IP address’ on page 12-43).
3. Save TCP/IP settings. The changes will be effective after the system is rebooted.
4. Network Settings: Configure system settings for network connection.
5. Video Streaming Settings: Configure streaming source for View-X (see page 14-13).
6. Wireless Settings: Configure WiFi network connection (see page 12-54).
7. Data Streaming Settings: Configure streaming of live ultrasound image data.