Editing/creating a Stress Echo protocol template
Vivid S70 / S60User Manual 7-29
BC092760-1EN 01
Assigning new labels to levels and projections
1. Select a label from the Label drop-down menu or type a new
label.
Configuring levels
The following options can be set up for each level:
Number of cycles to be stored in the cineloop
1. Enter the desired number in the Cycles field.
Up to four cycles/cineloop can be stored.
NOTE: Even if you store more than one cycle, while in analysis of
the Stress echo images the system will by default only
display the last cycle.
Continuous capture
1. Check Continuous capture if continuous image acquisition
throughout the level is desired.
When Continuous capture is selected, preview of cineloop
and reference display (see below) during acquisition are not
possible.
Preview of store
1. Check Preview of store if review and adjustment of
cineloops before storage is desired.
Show reference
1. Check Show reference if the display of the corresponding
reference loop is desired during acquisition (dual screen
mode).
Adding a group
1. In the Protocol template preview field select the cells to be
part of the group.
2. In the Pre-defined group field, press New group.
A dialogue box is displayed asking the user to enter a name
for the new group.
3. Enter the group name.
4. Press OK.
The new group is displayed in the Pre-defined group field.