Safety
2-6 Vivid S70 / S60User Manual
BC092760-1EN
01
Neonatal trans-cranial & spinal scanning
TI time Note
0.0–0.7 Unlimited Monitor TIC. MI>0.7 should be used with caution in the presence of
contrast agents.
0.7–1.0 < 60 min
1.0–1.5 < 30 min
1.5–2.0 < 15 min
2.0–2.5 < 4 min
2.5–3.0 < 1 min
Neonatal general and cardiac scanning
TI time Note
0.0–1.0 Unlimited Monitor TIB. MI>0.7 should be used with caution in the presence of
contrast agents.
1.0–1.5 < 120 min
1.5–2.0 < 60 min
2.0–2.5 < 15 min
2.5–3.0 < 4 min
Adult trans-cranial scanning
TI time Note
0.0–1.0 Unlimited Monitor TIC. MI>0.7 should be used with caution in the presence of
contrast agents.
1.0–1.5 < 30 min
1.5–2.0 < 15 min
2.0–2.5 < 4 min
2.5–3.0 < 1 min
General Abdominal, Peripheral Vascular and other scanning (excluding the eye)
TI time Note
0.0–1.0 Unlimited Monitor TIB or TIC if bone closer than 1 cm, TIS if no bone is in the
image. MI>0.7 should be used with caution in the presence of
contrast agents.
1.0–1.5 < 120 min
1.5–2.0 < 60 min
2.0–2.5 < 15 min
2.5–3.0 < 4 min