Customize Configuration
12-24 Vivid S70 / S60User Manual
BC092760-1EN
01
Creation of a user-defined folder
Figure 12-11. The Measurement menu sheet (Add folder)
1. Press Utility/Config on the Touch panel and select the
category Measure.
2. MCO and AV ET are Doppler measurements, select
Doppler in the Measurement menu sheet.
3. Select Add folder.
4. Give the folder a name (e.g. “My Folder”).
Adding measurements
Figure 12-12. The Measurement menu sheet (Add measurement)
1. Select the user-defined folder (e.g. “My Folder”) in the
Measurement menu.
2. Press Add Measurement in the Measurement menu sheet.
The Add measurement window is displayed.
1. Select the appropriate scanning mode.
2. Create a folder in the Measurement menu.
1. Select the user-defined folder.
2. Press Add measurement.