Probes
13-18 Vivid S70 / S60User Manual
BC092760-1EN
01
Table 13-2: Automated High-level Disinfectants Compatible with Non-TEE probes
Chemical Name Manufacturer
3Sc-RS
11L-D
12S-D
6S-D
9L-D
C1-6-D
C1-5-D
C2-9-D
C3-10-D
IC5-9-D
L8-18i-D
M5Sc-D
P2D
P6D
trophon EPR Nanosonics Limited X X X XXXXX XXX
trophon2 Nanosonics Limited X X X XXXXX XXX
Table 13-3: Description of Pictogram on Probe Care Cards
Pictogram Description Standard
“ATTENTION” - Consult accompanying documents” is
intended to alert the user to refer to the operator manual
or other instructions when complete information cannot be
provided on the label.
ISO 7000-0434A
“CAUTION” - Dangerous voltage (the lightning flash with
arrowhead) is used to indicate electric shock hazards.
IEC 60417-6042
Biohazard - Patient/user infection due to contaminated
equipment.
Usage
• Cleaning and care instructions
• Sheath and glove guidelines
ISO 7000-0659
Ultrasound probes are highly sensitive medical
instruments that can easily be damaged by improper
handling. Use care when handling and protect from
damage when not in use.
N/A- by GE Healthcare
Do not immerse the probe into any liquid beyond the level
specified for that probe. Refer to the user manual of the
ultrasound system.
N/A- by GE Healthcare
Since there is a possibility of having negative effects on
the probe, observe the specified immersing time by the
germicide manufacturer strictly. Do not immerse the probe
in liquid chemical germicides more than the time
prescribed in the care card.
N/A- by GE Healthcare
“Consult accompany document” - Refer to the ultrasound
system user manual for important probe care and cleaning
instruction.
ISO 7010-M002