Probes
13-26 Vivid S70 / S60User Manual
BC092760-1EN
01
Probe Intermediate-Level Disinfection - Wipe
1. Put on a new pair of gloves. Holding the probe near the
strain relief, apply the wipe to the patient contacting lens.
Wipe the probe from the lens to the strain relief, slightly
rotating the probe after each wiping pass.
2. After the probe has been completely wiped, use a second
wipe and starting at the probe lens begin wiping the probe in
a rotating motion moving down towards the strain relief.
Wring the wipe above recessed areas, seams, and ridges to
drip disinfectant directly onto these less accessible
surfaces.
Figure 13-10. Disinfecting the Probe Moving from Lens to
Strain Relief
3. Once the probe has been completely wiped, use a third
wipe and continue wiping the probe as needed to ensure the
surface remains wet for the required exposure time. Use as
many wipes as needed and drip additional disinfectant on
recessed areas and ridges, to ensure all surfaces remain
wet for the minimum required contact time listed in the
disinfectant manufacturer’s instructions for use.