Switching On/Off
Vivid S70 / S60 – User Manual 3-25
BC092760-1EN 01
NOTE: In case of total lockup of the system, hold the on/off button
down a few seconds to turn the system off.
2. Select Shutdown.
The shutdown process takes a few seconds and is
completed when the color of the on/off button changes from
green to amber.
3. Select Standby.
The systems enters Standby mode.
Booting up from Standby mode takes less than a minute.
NOTE: After switching off the system, wait at least ten seconds before
turning it on again.
NOTE: Whenever the system is fully shut down or put in Standby mode,
the system automatically performs "End-Exam" to save all data
and images of the current patient into the archiving system.
When switching the system off to move it, follow the additional
steps below:
1. Wait until the on/off button has turned amber and set the
circuit breaker to OFF.
2. Remove the plug from the mains power socket.
3. Secure the system power cable around the cable storage
hooks at the rear of the system.
Battery performance may degrade over time and limit the use
of transportation mode. In case the battery can not be fully
charged, or can not support transportation mode for more than
20 minutes please arrange for service of the battery.