EasyManua.ls Logo

GE Voluson E Series - Page 197

GE Voluson E Series
356 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.14.3.1 Auto Tester
Autotester is a log function of customer activities. It records all user actions (scanning, Touch Panel entries,
performing Calculations, review of Patient Reports, etc.…). It is possible to safe (record) as file on HDD. But
also export to DVD/CD can be done to allow replay of the records on other units.
Note
For intermittent problems this file can be requested from the Service Engineer or customer. It is possible to
burn this file on DVD/CD+R/RW.
Operation see
Section 7.7 "How to use the Auto Tester program" on page 7-18
.
5.14.3.2 Additional Software Installation (ASI)
Click the
ASI
button to install additional software. The Software parts to be installed depend on the contents
of the System DVD that is used.
5.14.3.3 FMI from DVD
By means of the
FMI from DVD
button, the Systems C:\ image is partly or completely updated. The System
Software parts to be upgraded depend on contents of the used System DVD.
The first “Boot Device” in BIOS has to be Hard Disk Drive.
Note
During “FMI from DVD” the used system configuration (incl. Full Backup) will be stored on R:\. If required, the
previously used System configuration (before FMI from DVD was performed) can be restored by activating
the “Rollback” function. Operation see: Section 5.15.3 on page 5-53 .
5.14.3.4 Restore Software
Click the
Restore software
button to perform an automatic restore of the system software from your hard
disk. The installation procedure starts with saving and recording the settings present on the system (silent
“Full Backup" and "Rollback”). During the software restore the system will restart several times.
Note
Please make sure that ONLY the DVD writer is connected on the USB ports. Disconnect all other external
USB devices (such as printers, hubs, bluetooth, memory devices) as this might interfere with the recovery/
installation procedure.
Note
Existing User Programs, 3D/4D Programs and Auto Text remain unaffected! Therefore it is not necessary to
perform any readout preparations.
5.14.3.5 Decrypt Disk
Note
Security of patient data on the system will be reduced when encryption is turned off. Decryption will take a
considerable amount of time (one to several hours).
5.14.3.6 TCP/IP Buffersize
The TCP/IP Buffersize selects the amount of buffer memory used for DICOM transfers (both directions).
5.14.3.7 Common Service Desktop (CSD)
Access to the Common Service Desktop (CSD) by entering security level and password. Each user is only
granted access to the tools that are authorized for their use.
5.14.3.8 Request for Service (RFS)
Fill out the “Request For Service” form and then send the problem description to GE Service/Application
representatives. Operation see
Section 7.2 "Request for Service (RFS)" on page 7-5
.
Components and Functions (Theory)
Voluson E-Series Service Manual
5539550APB Revision 6
5-51

Table of Contents