Operation Therapy
User Manual corpuls
3
92 ENG - Version 2.1 P/N 04130.2
The CPR feedback will be available in devices as of May 2013. Older devices
can be retrofitted. For further information, please contact your authorised sales
and service partner.
The corPatch CPR sensor is a disposable article.
The corPatch CPR sensor is covered under one or more of the following
U.S.A. patents: 7,074,199; 7,108,665; 7,429,250; 8,147,433; 7,220,235.
WARNING
To guarantee a defibrillation protection for patients, users and third parties,
exclusively use the accessories indicated in the list of authorised accessories.
Caution
The detection of the rate and depth of the thorax compressions can be
compromised by vibrations.
Caution
Do not use corPatch CPR sensors if
the packaging is damaged or opened;
the expiry date indicated on the packaging has passed;
the sensor is damaged.
Caution
Used corPatch CPR sensors have to be replace by a new corPatch CPR
sensor after 24 hours at the latest, to avoid side effects such as reddening
and skin irritation.
Caution
The patient's condition has to be assessed by the users themselves,
independent of the CPR feedback.
WARNING
During the phase between compressions, the user has to make sure that the
pressure is completely relieved from the thorax. Otherwise, there can be
false-negative feedback.
Note
Note
Note