User Manual corpuls
3
General Operating Instructions
ENG - Version 2.1 P/N 04130.2 43
4.2 Switching On and Off
4.2.1 Switching On
Press the On/Off key on the monitoring unit.
All modules are switched on.
Press the
AED
or
Man
ue
ll
Manual
key on the monitoring unit.
The corpuls
3
either starts in AED mode or in manual defibrillation mode,
respectively.
1. Press the On/Off key on the monitoring unit.
The monitoring unit and the module connected mechanically to it are
switched on.
2. Press the On/Off key on the module which is not connected mechanically.
This module is switched on.
Press the On/Off key on all individual modules.
The modules are switched on independently.
When a switched-off module is connected mechanically to a switched-on
module, it switches on automatically.
If a battery is inserted into a module, this module switches on automatically.
A certain amount of time is required to achieve operational status after
switching on. It is therefore recommended to switch on the corpuls
3
as soon
as possible.
Warning
If the alarm message „ß SW-NOT FOR PAT. USE“ appears after the device is
switched on, the software of one or all modules is a betaversion.
Using this version on a patient is prohibited. Please contact your sales and
service partner.
Warning
If the alarm message „ONLY FOR TEST PURPOSE“ appears after the device
is switched on, the software of one or all modules is a test version.
Using this version on a patient is prohibited. Please contact your sales and
service partner.