Configuration
User Manual corpuls
3
184 ENG - Version 2.1 – P/N 04130.2
7.4.14 Configuration of Non-Invasive Blood Pressure
Measurement (NIBP) (Persons Responsible for the
Device)
Persons responsible for the device can configure advanced settings for interval
settings, the initial mode and the initial pressure.
1. In the main menu, select "NIBP" ► "Settings".
The configuration dialogue opens.
Fig. 7-31 NIBP settings
2. Select the settings.
Table 7-21 shows the possible values
Field Setting Values Increment
Automatic Interval
1 min, 2 min, 3 min,
5 min, 10 min,
15 min,30 min,
60 min
–
Initial Mode Patient
Adult
- Child
Neonate
Initial Pressure
Adult 120 to 280 mmHg 10
Child 80 to 250 mmHg 10
Neonate 60 to 140 mmHg 10
7-21 Values for NIBP monitoring