Configuration
User Manual corpuls
3
166 ENG - Version 2.1 P/N 04130.2
The filter value of the high pass filter corresponds to the lower limit frequency of
the filter.
The filter value of the low pass filter corresponds to the upper limit frequency of
the filter.
2. Select the settings.
Table 7-14 contains the possible values.
Fields Settings Values
Monitoring
Low pass 25 Hz; 40 Hz; 150 Hz
High pass
0.5 Hz; 0.25 Hz;
0.12 Hz; 0.05 Hz
Diagnostic Low pass 40 Hz; 150 Hz
Table
7-14 Filter setting for monitoring ECG, diagnostic ECG (persons
responsible for the device)
For information pertaining to the configuration of ECG measurement and ECG
interpretation, see chapter 7.4.10 Configuration of ECG Measurement and ECG
Interpretation (Persons Responsible for the Device), page 179.
7.4.5 Alarm Configuration (Persons Responsible for the
Device)
In addition to the settings described in chapter 7.3.1 Configuring Alarm Settings,
p. 155, persons responsible for the device can configure the following settings
in the alarm configuration:
create an event in the protocol in case of alarms
configure the settings for alarm suspension (15-120 s or permanently)
Alarm settings
Note
Note