9.2.1 Function check of the Device
For a function check of the compact device, all the modules of the corpuls
3
must be connected
mechanically. The mechanical connections have to be heard to click into place. The following
measures must be taken:
Functional
test
Description Measures by the user Correct result
Switch-on Start of
functional test
Press the On/Off key
The start logo appears
Internal self
test
The corpuls
3
performs a test
of the internal
functions
None The jog dial briefly illuminates
at the beginning.
The display is illuminated.
Curve and parameter fields
are displayed.
The state of charge of the
batteries is displayed in
percent or in minutes
remaining running time.
Sufficient state of charge
present.
Connection status of the
modules is displayed.
Connection
test of the
modules
Function
check of the
network
Separation of all modules
Re-connecting all modules
In the status line the
connection status changes
from bar symbol to wave
symbol
In the status line the
connection status changes
from wave symbol to bar
symbol
No errror message is issued
Visual check
of the
corpuls
3
Detection of
alterations on
the corpuls
3
Check the complete
corpuls
3
for alterations
No objections
Defibrillator/
pacer
Visual check
of the device
and
accessories of
the
defibrillator/pa
cer
Remove shock paddles.
Completely unwind the
therapy master cable from
the socket.
Inspect shock paddles and
therapy master cable for
damage.
Check packaging of
corPatch electrodes,
verification of the expiry
date.
Check if replacement
electrodes are present.
Shock paddles and therapy
master cable are undamaged.
Packaging of corPatch
electrodes is undamaged and
replacement available.
Expiry date of corPatch
electrodes has not passed.
All in all, no damage is
detectable.