EasyManua.ls Logo

GS corpuls3 - F Safety Information

GS corpuls3
346 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Manual corpuls
3
Appendix
ENG - Version 2.1 P/N 04130.2 303
Precision of the Energy Released for Shock Spoons
Selected
energy
(in Joule)
Nominal energy released in comparison to patient
impedance
Accuracy:
Load impedance (in Ohm)
10
15
20
25
50
100
150
2
1.5
1.5
1.7
1.7
2.2
2.1
2.0
± 3J
5
4.0
4.2
4.5
4.9
5.6
5.5
5.2
± 3J
10
8.8
9.2
9.0
9.2
11.2
11
10.2
± 15 %
20
18.1
19.4
20.2
20.8
22.8
23.7
22.6
± 15 %
30
26.4
28.6
29.5
31.3
34.5
33.4
32.7
± 15 %
40
36.8
37.7
39.8
40.8
45.9
45.8
43.3
± 15 %
50
46.3
49.6
49.2
51.2
57.7
56.1
54.7
± 15 %
Table A-32 Precision of the energy released for shock spoons
F Safety Information
Warning
Failure to observe this safety information may result in injuries to the patient and
users.
General
The corpuls
3
is not intended for operation in the vicinity of readily inflammable anaesthetics or
other flammable substances, particularly in an oxygen-rich environment.
The corpuls
3
must not be stored or operated near a switched-on magnetic resonance imaging
(MRI) unit.
The corpuls
3
must not be operated near ionising (radioactive) radiation intended for therapeutic
purposes.
The corpuls
3
is not intended for operation in connection with a high frequency surgical device.
To avoid stumbling or the patient, user or third parties being entangled in or strangled by cables,
they should be led around the patient, as all medical implements.
When connecting intermediate cables and sensors, make sure there are no foreign objects
between the plug connections.
Alarm function
Do not leave the patient unattended in the following cases:
- if the alarm function is deactivated or
- device is in defibrillation mode.
In case of unsupervised monitoring of vital parameters, it is recommended that you monitor a
further vital parameter with another independent monitoring system.
The option for monitoring measured values may be disabled when switching on the corpuls
3
(see chapter 7 Configuration, p. 137). Therefore make sure that the alarms are correctly
configured.
Heart rate is only monitored when all electrodes of the ECG monitoring cable or the corPatch
electrodes are connected to the patient (see chapter 7 Configuration, p. 137).
Check the settings of the alarm limits before each monitoring phase.
Set the volume of the acoustic alarm so that it cannot be overheard in loud environments.

Table of Contents

Other manuals for GS corpuls3