Appendix
User Manual corpuls
3
310 ENG - Version 2.1 P/N 04130.2
H corpuls3 HYPERBARIC (HBO)
The product variant corpuls
3
HYPERBARIC is approved for operation in a
multiplace hyperbaric chamber during hyperbaric oxygen therapy (HBO) up to
an overpressure of 3 barg and a maximum oxygen concentration of 23%
(Certificate No. 77492-11, Germanischer Lloyd SE, Hamburg).
For further information that go beyond this appendix, please contact the
manufacturer or their authorised sales and service partners.
When the door of the hyperbaric chamber is closed, the reach of the wireless
connection towards modules outside the chamber is reduced.
Caution
There can be changes in the measured values of non-invasive blood pressure
measurement (option) during the build-up and decrease of pressure inside
the pressure chamber.
Do not perform NIBP measurements during the diving phases.
Caution
The can be changes in the measured values of CO
2
monitoring (option),
because the CO
2
partial pressures are changing under pressure.
Warning
Charging the monitoring unit and the patient box inside the pressure chamber
is only allowed via the charging bracket for monitoring unit, P/N, 04401.
The use of MagCode connectors and charging brackets with integrated
MagCode is forbidden for charging the monitoring unit, the patient box and
the defibrillator/pacer.
The red cover over the magnetic contact fields may not be removed.
Warning
When administering oxygen via inhalation mask or nasal cannula or when
using a ventilation bag with reservoir, make sure that no oxygen accumulates
above the corPatch electrodes. When defibrillation is imminent, the oxygen
supply has to be shut off.
Warning
The use of swivelling adapters (swivelling adapter 35 °, P/N 04406.01 and
swivelling adapter 60 °, P/N 04406) is forbidden inside the pressure chamber.
Warning
The protection of the power supply for the charging bracket for the monitoring
unit, P/N 04401 has to be minimum 6 A and maximum 10 A.
Warning
To guarantee safe operation of the device in a hyperbaric chamber,
exclusively use the materials itemised in the list of accessories approved for
operation in hyperbaric chamber (see chapter 9.8 Approved Accessories,
Spare Parts and Consumables, page 224).
Modules, batteries and accessories without HBO certificate may not be used
in combination with HBO certified modules, batteries and accessories.
Operation in
hyperbaric chamber
up to 3 barg and
23% oxygen