Procedure in case of malfunctions
ENG - Version 2.1 – P/N 04130.2 233
10 Procedure in Case of Malfunctions
10.1 Device alarms
The following table lists all alarms of the device with their priority and describes the cause of the
malfunction as well as how to eliminate it.
Alarm message Priority Explanation/Measure
Battery low (Defib)
Battery low (Monitor)
Battery low (P-Box)
High priority The state of charge of the battery is now less than
20% of the total charge of the module indicated.
Connect the respective module to a power supply as
soon as possible.
Battery temperature
high
(Defib)
Battery temp. high
(Monitor)
Battery temp. high
(P-Box)
Medium priority The battery charging process lead to a great
increase of temperature inside the battery (> 60°C).
The corpuls
3
or the respective module was
possibly exposed to high temperatures. See
appendix D Technische Daten, page 5.
Interrupt charging process by disconnecting the
corpuls
3
from the mains power or from charger.
If necessary, let the corpuls
3
cool down or bring
device to a cooler environment.
Contact authorised sales and service partners, if
necessary.
BT connected:
[DEVICE]
Medium priority The connection to another Bluetooth
®
device has
been established.
"Device" stands for a ventilator (e. g. Weinmann
Medumat Transport) or another input device (e. g. a
tablet PC).
BT connection failed Medium priority The connection to another Bluetooth
®
device has
failed.
Check the Bluetooth
®
connections.
BT disconnected:
[DEVICE]
Medium priority The connection to another Bluetooth
®
device has
been interrupted.
"Device" stands for a ventilator (e. g. Weinmann
Medumat Transport) or another input device (e. g. a
tablet PC).