Procedure in case of malfunctions
User Manual corpuls
3
234 ENG - Version 2.1 P/N 04130.2
Alarm message Priority Explanation/Measure
CF card almost full Low priority The current capacity of the memory card
(CompactFlash® card) amounts to less than 20% of
the memory capacity or the number of missions is
more than 999.
Remove the memory card (CompactFlash
®
card)
and save the contents on another storage medium
(e.g. PC).
Delete the data at regular intervals once they are no
longer required.
If the CF card is full, the mission data are no longer
recorded.
CF card error Low priority The memory card (CompactFlash® card) is not
correctly formatted. Remove the memory card and
back up the contents on other storage media (e.g.
PC).
Format according to chapter 8.3 Handling Data,
page 187.
The memory card (CompactFlash
®
card) is faulty.
The mission data of the current or of future missions
are no longer recorded.
CF card full
Low priority Remove the memory card (CompactFlash
®
card)
and save the contents on another storage medium
(e.g. PC).
Delete the data at regular intervals once they are no
longer required.
The mission data of the current or of future missions
are no longer recorded.
CF card missing Low priority The memory card (CompactFlash
®
card) is not
correctly inserted in the drawer of the patient box.
The memory card (CompactFlash
®
card) is missing.
The mission data of the current or of future missions
are no longer recorded.
Check battery (Defib)
Check battery (Monitor)
Check battery (P-Box)
Medium priority The battery has to be checked as soon as possible
and replaced, if necessary.
Contact authorised sales and service partners.