ENG - Version 2.1 P/N 04130.2
2 Intended Use
The corpuls
3
is intended
for measurement and monitoring of vital functions in addition to
defibrillation, cardioversion or cardiac pacing
of patients in the preclinical and clinical field by qualified medical staff trained in
the use of the device.
The following monitoring functions are available:
ECG
diagnostic ECG
Optional:
oximetry (SpO
2,
SpCO
®,
SpHb, SpMet
®
)
capnometry (CO
2
)
temperature (T)
non-invasive blood pressure monitoring (NIBP)
invasive blood pressure monitoring (IBP)
The corpuls
3
is approved for monitoring in operating diagnostic X-ray units
(e. g. computed tomography). Exempt from this is the oximetry option, because
measured values might be falsified. When equipped with the HBO (hyperbaric
oxygen therapy) option, the corpuls
3
is approved for operation in a multiplace
hyperbaric chamber up to 3 barg and an oxygen concentration of < 23%.
Intended use of corpuls
3
includes employment of accessories which are
approved by the manufacturer (see chapter 9.8 Approved Accessories,
Spare Parts and Consumables, p. 224) and
appropriate for the function and patient.
Use of accessories on corpuls
3
which are not approved by the manufacturer
is not considered to be intended use.
Warning
Defibrillation protection for patients, user and third parties cannot be
guaranteed, if accessories other than those authorised by the manufacturer
are used.
The therapeutic functions of defibrillation, cardioversion and pacer must only be
performed with constant monitoring of the patient.
Performing the therapeutic functions without eye contact with the patient is not
considered to be intended use.
Intended use