User Manual corpuls
3
Configuration
ENG - Version 2.1 – P/N 04130.2 183
7.4.13 Configuration of the Metronome (Persons
Responsible for the Device)
Persons responsible for the device can select advanced settings for the
metronome and for the CPR feedback system in order to adapt to local
variations of resuscitation algorithms. For this purpose, the
• Compression frequency and the
• Duration of the ventilation phases
can be configured for adults and children.
The person responsible for the device can adapt the resuscitation protocol to
locally applicable variations. Besides the currently valid AHA/ERC 2010
guidelines, a 3 min-resuscitation protocol is available.
For configuration of the metronome, proceed as follows:
1. In the main menu, select "Defib" ► "CPR".
2. Select the required metronome setting with the jog dial and confirm.
3. Select the required setting and confirm with softkey [OK].
4. Save configuration (see chapter 7.4.2 General System Settings (Person
responsible for the device), page 160).
Field Setting Values
Adult
Compress.
Compression
frequency
80 – 120 1/min
increments of 5
1/min
Vent. 30:2
Duration of
ventilation at ratio
30:2
3 – 6 s
increments of 1 s
Child
Compress.
Compression
frequency
80 – 140 1/min
increments of 5
1/min
Vent. 15:2
Duration of
ventilation at ratio
15:2
3 – 8 s
increments of 1 s
Vent. 30:2
Duration of
ventilation at ratio
30:2
3 – 8 s
increments of 1 s
Algorithm Interval
Duration of
resuscitation
protocol
2 min; 3 min
7-20 Values for metronome settings
Advanced
Metronome
Settings