Maintenance and tests
ENG - Version 2.1 – P/N 04130.2 213
9.2.2 Function check of the Power Supply
Check Description Measures by the user Correct result
Batteries for
individual
modules
Check for presence Check if each module of the
corpuls
3
has a battery
inserted.
All modules of corpuls
3
have an inserted battery.
State of
charge of the
battery
Check the state of
charge of the
corpuls
3
batteries
as a compact
device
Connect all modules of the
corpuls
3
(compact
device).
Connect corpuls
3
to the
mains supply (either via
the charging bracket or
and external mains
charger).
Switch on the corpuls
3
and check the state of
charge of the batteries
(percentage) in the status
line after the booting
process.
• The state of charge of the
batteries shows a figure
greater than 30%.
• If the corpuls
3
will
probably be operated at
low temperatures (e.g. in
the cold season) the state
of charge should indicate
a figure higher than 50%
(at room temperature)
Remaining
running time
Check the
remaining running
time of the
individual modules
Separate all corpuls
3
modules in the switched-
on state.
Remove the charging
cable of the mains charger
or remove the module
from its charging bracket.
Check the predicted
remaining running time of
the monitoring unit and
patient box in the status
line.
The remaining running time
of the individual modules is
collectively more than
120 minutes.
9-3 Function check of the power supply
9.2.3 Checking Accessories and Consumables
Check Description Measures by the user Correct result
corPatch
electrodes
Check presence
and serviceability of
corPatch
electrodes for
adults and
neonates
Check if at least two pairs
of serviceable corPatch
electrodes are present.
Check if the packaging of
the corPatch electrodes
is undamaged.
Check if the expiry date of
the corPatch electrodes
has passed.
At least two pairs of
corPatch electrodes are
present.
The packaging of the
corPatch electrodes is
undamaged.
The expiry date has not
passed.