Procedure in case of malfunctions
ENG - Version 2.1 – P/N 04130.2 263
10.3 Notifications Message Line and Information in the Protocol
The notifications marked by `- -` do not require further explanation, as they are self-explanatory. The
measure to be taken is to follow the instruction given in the notification. The following notification texts
can be shown on the display and/or are printed out in the protocol (in alphabetical order*):
message line and
information in the
protocol
Event in the protocol, recording that for the vital parameter no value
has yet been measured.
Event in the protocol, recording that the vital parameters could not
be documented.
[NUMBER] ?
Event in the protocol, recording that the issued vital parameter has
exceeded or fallen below an alarm limit.
[NUMBER] ?
Event in the protocol, recording that the issued vital parameter is
not reliable.
[NUMBER] ?
Event in the protocol, recording that the issued vital parameter is
not reliable but also has exceeded or fallen below an alarm limit.
Event in the protocol that the defibrillator has been activated in AED
mode.
Event in the protocol that the defibrillator has been activated in
manual mode.
[MODULE]
The Ad-hoc connection with the patient box or the defibrillator/pacer
has been established.
[NUMBER] sec
Message in the status line that the alarm suspension is activated. The
count-down shows the remaining time in seconds until the alarm
suspension ends automatically.
Message in the message line after the connection to another
Bluetooth
®
device has been established.
"Device" stands for a ventilator (e. g. Weinmann Medumat
Transport) or another input device (e. g. a tablet PC).
Message in the message line after the connection to another
Bluetooth
®
device has failed.
Check the Bluetooth
®
connections.
BT disconnected: [DEVICE]
Message in the message line after the connection to another
Bluetooth
®
device has been interrupted.
"Device" stands for a ventilator (e. g. Weinmann Medumat
Transport) or another input device (e. g. a tablet PC).
available
Currently no data can be saved on or read from the CompactFlash
®
card.
Check if a CompactFlash
®
card is inserted.
In modular operation: reduce the distance between the monitoring
unit and the patient box, if necessary.
Connect the modules, if necessary.